Your activity: 34112 p.v.
your limit has been reached. plz Donate us to allow your ip full access, Email: [email protected]

Percutaneous pulmonic valve implantation

Percutaneous pulmonic valve implantation
Author:
Andreas Eicken, MD, PhD, FESC
Section Editor:
Heidi M Connolly, MD, FACC, FASE
Deputy Editor:
Susan B Yeon, MD, JD, FACC
Literature review current through: Feb 2022. | This topic last updated: Sep 02, 2020.

INTRODUCTION — Percutaneous pulmonic valve implantation (PPVI) was developed as a nonsurgical treatment for patients with right ventricular outflow tract (RVOT) dysfunction [1]. PPVI devices are intended for use in a dysfunctional (stenotic or regurgitant) right ventricle-to-pulmonary artery conduit. PPVI is intended to extend the lifetime of a right ventricle-to-pulmonary artery conduit and hence reduce the total number of open-heart surgeries required over a patient’s lifetime.

Some centers have extended the application of PPVI to treat patients with tetralogy of Fallot without a conduit [2,3] and to treat patients with tetralogy of Fallot with failing pulmonic bioprosthetic valves [4,5].

INDICATIONS AND EXCLUSIONS

Our approach — We recommend PPVI for patients with right ventricle to pulmonary artery conduits, native right ventricle outflow tracts, or failing bioprosthetic valves in the pulmonary position meeting the following criteria for severe right ventricular outflow tract (RVOT) obstruction or severe pulmonic regurgitation [3,4,6-8]:

Severe RVOT obstruction (as defined in the bullets below) with no or mild pulmonary regurgitation:

With normal RV systolic function

-Symptoms (<65 percent of expected exercise tolerance) related to RVOT obstruction plus a peak continuous wave Doppler velocity of >3.5 m/s.

or

-No symptoms plus RVOT obstruction with peak continuous wave Doppler velocity >4.0 m/s or RVOT obstruction with right ventricular systolic pressure greater than two-thirds of systemic systolic pressure.

With RV systolic dysfunction – Generally similar thresholds are used as with normal RV systolic function. However, some patients with severe RV systolic dysfunction may have severe RVOT obstruction with gradients that do not meet the standard thresholds for intervention due to low cardiac output; some of these patients may benefit from PPVI, although data are lacking to support this approach.

Severe pulmonary regurgitation and little or no RVOT obstruction, with either symptoms, arrhythmias, or right ventricular end-diastolic volume index >150 cm/m² by cardiovascular magnetic resonance imaging.

Candidates for PPVI must also meet the following two criteria:

Adequate RVOT conduit size to accommodate a covered stent. We do not have a lower size limitation. The upper size limitation is usually a diameter <29 mm. In selected cases, a slightly larger diameter may be successfully treated after pre-stenting to a smaller diameter to create a suitable "landing zone" or by overdilating an Edwards SAPIEN valve 29 to 31 mm.

Adequate body size. There is no absolute lower age limit but an adequate body size (eg, weight >20 kg) is required to accommodate femoral placement of the introducer [7].

The following exclusion criteria are adapted from the 2010 European Society of Cardiology (ESC) guidelines for the management of grown-up congenital heart disease endorsed by the Association for European Pediatric Cardiology (AEPC) [8,9]:

Evidence of risk of coronary compression by the expanded implant as determined by balloon testing (image 1) [10].

Central vein occlusion or significant obstruction.

Active infection (such as endocarditis) or high risk of infection (such as intravenous drug abuse).

Surgery is preferred when additional interventions are considered such as tricuspid annuloplasty, coronary artery bypass, or arrhythmia surgery.

Major society guidelines — The 2010 ESC/AEPC guidelines for the management of grown-up congenital heart disease include the following indications for surgical intervention or PPVI in patients with right ventricular to pulmonary artery conduits [9]:

Intervention is recommended in symptomatic patients with RV systolic pressure >60 mmHg (TR velocity >3.5 m/s; may be lower in cases with reduced flow) and/or moderate to severe pulmonic regurgitation.

Intervention is suggested in asymptomatic patients with severe RVOT obstruction and/or severe pulmonic regurgitation when at least one of the following criteria is present:

Decrease in exercise capacity on cardiopulmonary exercise testing

Progressive RV dilation

Progressive RV systolic dysfunction

Progressive TR (at least moderate)

RV systolic pressure >80 mmHg (TR velocity >4.3 m/s)

Sustained atrial/ventricular arrhythmias

IMPLANTATION TECHNIQUE — Two different types of valves are commercially available for PPVI, the Medtronic Melody Transcatheter Pulmonary Valve and the Edwards SAPIEN Pulmonic Transcatheter Heart Valve. Initial results of a third self-expandable valve for right ventricular outflow tract (RVOT)-dysfunction are available [11,12].

The implantation technique for the two valves for PPVI is similar [1,7,10,13-16]. PPVI is usually performed using general endotracheal anesthesia or conscious sedation. Venous access is generally obtained through the femoral vein, although a jugular venous approach is also feasible. After a complete hemodynamic and angiographic study, a guidewire is positioned distally into a pulmonary artery.

Balloon test — Because coronary compression is a serious complication of PPVI, a balloon test to exclude potential coronary artery compression should always be performed. This involves inflation of a balloon in the RVOT conduit. During balloon inflation, an aortogram is obtained to display the coronary arteries (image 1) [10]. If coronary artery flow is impaired during this test, PPVI should not be performed. For this test, an aortogram may be preferable to selective coronary artery angiogram because intubation of the coronary artery by the catheter may lead to a false negative test. Initially, the balloon test is done with a low pressure balloon to prevent the risk of conduit rupture. The balloon diameter equals the intended external diameter of the landing zone for the pulmonic valve. In case of close proximity of the coronary arteries, a balloon test with a high pressure balloon (ie, Atlas Kevlar balloons) may be indicated as well as selective coronary angiograms in multiple planes.

Pre-stenting — Observational studies have found that pre-stenting of the conduit before PPVI is associated with a lower risk of developing PPVI stent fractures with the Melody valve [17]. Therefore, pre-stenting of the conduit before PPVI is now standard practice in most centers. Usually, a covered stent is preferred to avoid bleeding out of the ruptured conduit during deployment of the stent, because high-pressure balloons are needed to dilate the conduit to the previously tested diameter. In some cases (eg, long stenosis or significant recoil during balloon deflation), additional stents are needed for pre-stenting to support the landing zone. A non-covered stent is usually used if obstruction of the origin of one of the pulmonary branch arteries ("jailing") is expected after deployment of the pre-stent.

Valve types — Depending on the final diameter of the treated RVOT, either a Melody valve (balloon diameter 18, 20, or 22 mm and external valve diameter 20 to 24 mm) or a SAPIEN valve (23, 26, or 29 mm) is chosen.

The Medtronic Melody Transcatheter Pulmonary Valve (image 2) is available in two sizes (20 and 22). The 20 mm valve can be implanted with a diameter of 18 or 20 mm balloon and the 22 mm valve with an 18, 20, or 22 mm delivery system. The Melody valve consists of a 34 mm bare metal Cheatham platinum stent (NuMED CP Stent CP8Z34) into which a Medtronic Contegra bovine jugular venous valve is hand-sewn. The valve is crimped onto a designated double balloon delivery system requiring a 22F sheath.

The Edwards SAPIEN Pulmonic Transcatheter Heart Valve (THV) (image 3) is available in three sizes, 23, 26, and 29 mm. The SAPIEN valve is made of a trileaflet bovine pericardial valve hand-sewn into a stainless-steel stent. There is a fabric-sealing cuff covering the proximal part of the stent designed to prevent paravalvular leaking. The valve is crimped onto a designated balloon delivery system requiring a 14 or 16F sheath for the 23, 26, or 29 mm valve. The SAPIEN THV is advanced without protection into the RVOT. If the THV struts get caught in the tricuspid valve, the valve may be damaged [18]. To avoid this complication, some have advocated use of a large sheath for direct implantation of the valve in the RVOT [19].

Valve-in-valve — PPVI can also be performed as a repeat intervention after failure of the primary device in the RVOT [20]. The indications and the implantation technique are the same for repeat PPVI as for primary PPVI. Valve-in-valve technology has also been used in other prosthetic valve sites [21].

OUTCOMES — Short- and medium-term results of PPVI with the Melody and SAPIEN valves are similar, though more data are available for the Melody valve. Long-term data are available for the Melody valve, as discussed below. Long-term data are not yet available for the SAPIEN valve.

Melody valve — Data are available from four short- and medium-term observational studies with a total of over 450 patients with one- to five-year follow-up [7,8,22,23] as well as from longer-term studies. These studies excluded 4 to 8 percent of the patients planned for PPVI due to an unfavorable right ventricular outflow tract (RVOT) dimension or coronary artery compression.

After successful PPVI, the following findings were noted:

Overall mortality was 0 to 5 percent and was, as far as could be assessed, not related to the PPVI.

The device was unstable in 1 percent, a conduit rupture requiring surgery occurred in 0.8 to 2 percent, compression of the left main coronary artery after device deployment occurred in 0 to 1 percent, and obstruction of the origin of one of the pulmonary branch arteries ("jailing") in 0.6 percent.

The gradient over the RVOT fell significantly from median values of 35 to 40 to 10 to 20 mmHg. Pulmonary regurgitation decreased from median values of 16 to 27 percent to 1 to 2 percent. Median peak velocity over the RVOT was 1.9 to 2.7 m/s at one-year echocardiographic follow-up.

Freedom from Melody valve dysfunction or reintervention was 94 to 95 percent at one year. Patients not requiring reintervention had mild or less pulmonic valve regurgitation at one-year follow-up. Overall, surgery was required in 3 percent during follow-up with freedom from reoperation of 86 percent at 30 months.

Endocarditis was diagnosed in 1 to 3 percent of the cases up to five years after PPVI with Melody valves. (See 'Endocarditis' below.)

At six months, New York Heart Association (NYHA) class improved in 77 percent of the patients and worsened only in 0.8 percent. RV end-diastolic volume and RV mass decreased significantly as demonstrated by cardiovascular magnetic resonance imaging.

Stent fractures occurred in 20 to 21 percent in a series without pre-stenting and in 5 to 16 percent in a series with pre-stenting.

In an early study, the first 50 patients of the series and those with residual gradients >25 mmHg had a higher risk of reoperations compared with the rest of the study group [22].

Four long-term studies on PPVI with the Melody valve have been published [24-27]. The best long-term results (estimated event-free survival rate of 88 percent at 10 years) were observed among patients with a RVOT residual peak invasive gradient of <15 mmHg after the intervention. One study compared the long-term outcome of PPVI (Melody valve) with surgical bioprosthetic pulmonary valve replacement [28]. At 10 years’ follow-up, there was no difference in freedom from valve replacement between the groups.

Some data suggest that PPVI in patients with postoperative right ventricular outflow obstruction or pulmonic regurgitation and concomitant moderate to severe tricuspid regurgitation results in acute reduction in the severity of tricuspid regurgitation and improvement in right ventricular remodeling [29,30]. More data are needed on the impact of PPVI on tricuspid regurgitation.

SAPIEN valve — Comparison studies with conventional surgery are lacking. Data are available from two short- and medium-term observational studies with a total of 43 patients [15,16]. In the larger series of 36 patients, successful valve deployment was achieved in 33 of 34 attempts [16]. Pullback peak-to-peak systolic gradient across the conduit decreased from 27 to 12 mmHg. At six months’ follow-up, the number of patients in NYHA functional class I increased from 5 at baseline to 27. Pulmonic regurgitation was moderate or less in 97 percent of patients. One patient was treated with a second valve due to conduit-induced distortion of the initial implant.

In the COMPASSION study, a total of 69 patients (SAPIEN THV 23 [n = 22], SAPIEN THV 26 [n = 47]) were successfully and effectively treated, and at three years’ follow-up, 90 percent of the patients had only mild or less pulmonary regurgitation [31]. Freedom from reintervention was 94 percent, and freedom from infective endocarditis was 97 percent. Additionally, two studies reported off-label PPVI with the SAPIEN S3 valve with short follow-up [32,33].

COMPLICATIONS

Coronary artery compression — Compression of the left main coronary artery after device deployment occurs in approximately 0 to 1 percent of all cases of PPVI and nearly always has a fatal outcome [7,8,22,23]. Therefore, a balloon test to exclude potential coronary artery compression should always be performed prior to deployment of the valve [10]. A balloon test positive for coronary artery compression was initially identified in approximately 1 percent of cases referred to PPVI [22]. A later study found coronary compression during the balloon test in 5 percent of the patients [34]. The intervention was abandoned in these cases. (See 'Balloon test' above.)

Rarely, hematoma formation from conduit laceration may cause acute extrinsic coronary artery compression [35].

Conduit rupture — Contained conduit rupture may easily be treated with covered stent implantation. Noncontained conduit rupture is an emergency. If not controlled by covered stent implantation, emergency surgery may be necessary (less than 1 percent of patients after PPVI) [7,8,22]. Risk factors for conduit rupture include heavy calcification of the conduit and a right ventricular outflow tract (RVOT) homograft valve [36].

Stent fracture — Risk factors for stent fracture include PPVI without pre-stenting (particularly with the Melody valve), severely obstructed conduits, PPVI into a native outflow tract, and presence of PPV recoil during balloon deflation [37,38].

Stent fractures (image 4) occur in 20 to 21 percent with the Melody valve without pre-stenting [8,22] and in 5 to 16 percent with pre-stenting [7,23]. In the majority of cases, these stent fractures have no significant hemodynamic effect [38]. However, stent fractures may ultimately lead to embolization of stent parts or restenosis [22]. Observational data suggest that pre-stenting reduces the risk of stent fractures [17,37]. Therefore, pre-stenting of the conduit with a bare metal stent before PPVI is now standard practice in most centers. (See 'Pre-stenting' above.)

So far, stent fractures have not been reported after PPVI with the SAPIEN valve. However, in the COMPASSION study, 91 percent of the patients were pre-stented preceding SAPIEN implantation [31]. SAPIEN PPVI without pre-stenting has been performed, as illustrated by a report of 57 patients with no stent fracture observed with median 5.7 months follow-up [39]. The rationale for this approach is that the stent of the SAPIEN valve is stronger than the stent of the Melody valve and thus may be less prone to stent fracture.

Endocarditis — All bioprosthetic valves predispose the patient to infective endocarditis (IE). IE after PPVI has been observed in approximately 1 to 3 percent of patients during one- to four-year follow-up [7,8,22,23,40-43]. The largest investigation of IE following PPVI with the Melody valve included 311 patients pooled from three prospectively identified cohorts and followed for a median of 2.5 years [41]. The data were updated to a median follow-up of 5.1 years. The annualized incidence rate of IE was 3.1 percent per patient year, and the transcatheter pulmonary valve-related IE rate was 2.4 percent. At five years post-PPVI, freedom from IE was 89 percent. By multivariate analysis, age ≤12 years at implant and immediate post-implant gradient ≥15 mmHg were associated with development of IE [44].

The rate of IE after PPVI is similar to the generally observed rate of IE after surgical RVOT reconstruction, which ranges from approximately 0 to 5 percent of patients during one- to four-year follow-up [45-52]. However, the rate of IE after PPVI may be higher than that seen after RVOT reconstruction with a homograft. A retrospective study included 677 patients with 738 right ventricular outflow conduits implanted between 1989 and 2013 [53]. Kaplan-Meier survival free of IE for patients with surgically implanted homografts was 98.7 percent at five years and 97.3 percent at 10 years; for surgically implanted Contegra conduits, it was 87.8 percent at five years and 77.3 percent at 10 years; and for transcatheter Melody valves, it was 84.9 percent at five years. Thus, the Contegra conduit and Melody valved stents (both produced from bovine jugular vein) were associated with significantly higher incidence of IE than homografts. However, it should be noted that a Melody valve is often implanted in a stenotic homograft. Therefore, it may be that a relevant number of Melody valves were implanted in a homograft that already suffered from clinically silent IE. Thus, further studies are needed to clarify these important observations.

A single-center series reported a significantly higher rate of IE after PPVI with the Melody valve compared with the SAPIEN valve. At a median of 1.8-year follow-up, there were eight cases of IE among 32 (5.7 percent) patients who received a Melody valve and no cases of IE among 47 patients with a SAPIEN valve [54]. Since long-term data for the SAPIEN valve are not available, additional data are needed to confirm whether IE rates differ among valve types. In these comparisons, not only valve type but also indication for treatment (stenosis or regurgitation) needs to be included.

Vegetations on a transcatheter pulmonic valve or conduit can be difficult to visualize by echocardiography [41,55]. A report suggested that intracardiac echocardiography can be helpful when transthoracic and transesophageal echocardiography fail to show vegetations [42]. Alternate imaging with cardiac computed tomography, magnetic resonance imaging, or positron emission tomography may also be helpful.

All patients with prosthetic valves, including those who have undergone PPVI are considered among those at highest risk for IE and therefore prophylaxis for bacterial endocarditis is suggested for high risk procedures such as all dental procedures that involve manipulation of gingival tissue, the periapical region of teeth, or perforation of the oral mucosa. Recommendations for antibiotic prophylaxis for endocarditis are discussed in detail separately. (See "Antimicrobial prophylaxis for the prevention of bacterial endocarditis".)

Investigational approaches — Three self-expanding valves are currently being evaluated for approval. The Harmony transcatheter pulmonary valve, the venous P valve, and the Pulsta valve may enable treatment of patients with larger RVOTs [56-58].

Another approach is the implantation of an adaptive pre-stent to create a landing zone for secondary PPVI with the SAPIEN valve [59].

SOCIETY GUIDELINE LINKS — Links to society and government-sponsored guidelines from selected countries and regions around the world are provided separately. (See "Society guideline links: Cardiac valve disease".)

SUMMARY AND RECOMMENDATIONS

Percutaneous pulmonic valve implantation (PPVI) is an effective treatment for selected patients with right ventricular outflow tract (RVOT) conduit (or native) stenosis or regurgitation. Short- and mid-term results are satisfactory, long-term results are only available for the Melody valve, and direct comparison with conventional surgery is not yet available. (See 'Outcomes' above.)

We recommend PPVI in selected patients with severe RVOT obstruction or severe pulmonary regurgitation. (See 'Our approach' above.)

The most common complication following PPVI is stent fracture which usually is hemodynamically benign. Uncommon complications occurring in 2 percent or less of patients include clinically significant conduit rupture, coronary artery compression, and endocarditis. (See 'Complications' above.)

Since coronary compression is a life-threatening complication of PPVI, a balloon test is performed in the RVOT to exclude potential coronary artery compression (image 1). (See 'Implantation technique' above and 'Coronary artery compression' above.)

Melody valve stents seem to be associated with higher incidence of infective endocarditis than homografts, and vegetations may be challenging to visualize by transthoracic and transesophageal echocardiography. (See 'Endocarditis' above.)

ACKNOWLEDGMENT — The UpToDate editorial staff would like to thank Sohrab Fratz, MD, PhD, who contributed to earlier versions of this topic review.

REFERENCES

  1. Bonhoeffer P, Boudjemline Y, Saliba Z, et al. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction. Lancet 2000; 356:1403.
  2. Momenah TS, El Oakley R, Al Najashi K, et al. Extended application of percutaneous pulmonary valve implantation. J Am Coll Cardiol 2009; 53:1859.
  3. Boudjemline Y, Brugada G, Van-Aerschot I, et al. Outcomes and safety of transcatheter pulmonary valve replacement in patients with large patched right ventricular outflow tracts. Arch Cardiovasc Dis 2012; 105:404.
  4. Gillespie MJ, Rome JJ, Levi DS, et al. Melody valve implant within failed bioprosthetic valves in the pulmonary position: a multicenter experience. Circ Cardiovasc Interv 2012; 5:862.
  5. Tanase D, Grohmann J, Schubert S, et al. Cracking the ring of Edwards Perimount bioprosthesis with ultrahigh pressure balloons prior to transcatheter valve in valve implantation. Int J Cardiol 2014; 176:1048.
  6. Eicken A, Hess J. Percutaneous pulmonary valve implantation: the Munich experience. Interventional Cardiology 2012; 4:193.
  7. Eicken A, Ewert P, Hager A, et al. Percutaneous pulmonary valve implantation: two-centre experience with more than 100 patients. Eur Heart J 2011; 32:1260.
  8. McElhinney DB, Hellenbrand WE, Zahn EM, et al. Short- and medium-term outcomes after transcatheter pulmonary valve placement in the expanded multicenter US melody valve trial. Circulation 2010; 122:507.
  9. Baumgartner H, Bonhoeffer P, De Groot NM, et al. ESC Guidelines for the management of grown-up congenital heart disease (new version 2010). Eur Heart J 2010; 31:2915.
  10. Sridharan S, Coats L, Khambadkone S, et al. Images in cardiovascular medicine. Transcatheter right ventricular outflow tract intervention: the risk to the coronary circulation. Circulation 2006; 113:e934.
  11. Cao QL, Kenny D, Zhou D, et al. Early clinical experience with a novel self-expanding percutaneous stent-valve in the native right ventricular outflow tract. Catheter Cardiovasc Interv 2014; 84:1131.
  12. Husain J, Praichasilchai P, Gilbert Y, et al. Early European experience with the Venus P-valve®: filling the gap in percutaneous pulmonary valve implantation. EuroIntervention 2016; 12:e643.
  13. Zahn EM, Hellenbrand WE, Lock JE, McElhinney DB. Implantation of the melody transcatheter pulmonary valve in patients with a dysfunctional right ventricular outflow tract conduit early results from the u.s. Clinical trial. J Am Coll Cardiol 2009; 54:1722.
  14. Garay F, Webb J, Hijazi ZM. Percutaneous replacement of pulmonary valve using the Edwards-Cribier percutaneous heart valve: first report in a human patient. Catheter Cardiovasc Interv 2006; 67:659.
  15. Boone RH, Webb JG, Horlick E, et al. Transcatheter pulmonary valve implantation using the Edwards SAPIEN transcatheter heart valve. Catheter Cardiovasc Interv 2010; 75:286.
  16. Kenny D, Hijazi ZM, Kar S, et al. Percutaneous implantation of the Edwards SAPIEN transcatheter heart valve for conduit failure in the pulmonary position: early phase 1 results from an international multicenter clinical trial. J Am Coll Cardiol 2011; 58:2248.
  17. Nordmeyer J, Lurz P, Khambadkone S, et al. Pre-stenting with a bare metal stent before percutaneous pulmonary valve implantation: acute and 1-year outcomes. Heart 2011; 97:118.
  18. Faccini A, Butera G. Tricuspid regurgitation as a complication of Edwards Sapien XT valve implantation in pulmonary position a problem to deal with. Catheter Cardiovasc Interv 2018; 91:927.
  19. Kenny D, Morgan GJ, Murphy M, et al. Use of 65 cm large caliber Dryseal sheaths to facilitate delivery of the Edwards SAPIEN valve to dysfunctional right ventricular outflow tracts. Catheter Cardiovasc Interv 2019; 94:409.
  20. Nordmeyer J, Coats L, Lurz P, et al. Percutaneous pulmonary valve-in-valve implantation: a successful treatment concept for early device failure. Eur Heart J 2008; 29:810.
  21. Cullen MW, Cabalka AK, Alli OO, et al. Transvenous, antegrade Melody valve-in-valve implantation for bioprosthetic mitral and tricuspid valve dysfunction: a case series in children and adults. JACC Cardiovasc Interv 2013; 6:598.
  22. Lurz P, Coats L, Khambadkone S, et al. Percutaneous pulmonary valve implantation: impact of evolving technology and learning curve on clinical outcome. Circulation 2008; 117:1964.
  23. Butera G, Milanesi O, Spadoni I, et al. Melody transcatheter pulmonary valve implantation. Results from the registry of the Italian Society of Pediatric Cardiology. Catheter Cardiovasc Interv 2013; 81:310.
  24. Cheatham JP, Hellenbrand WE, Zahn EM, et al. Clinical and hemodynamic outcomes up to 7 years after transcatheter pulmonary valve replacement in the US melody valve investigational device exemption trial. Circulation 2015; 131:1960.
  25. Cools B, Brown S, Budts W, et al. Up to 11 years of experience with the Melody valved stent in the right ventricular outflow tract. EuroIntervention 2018; 14:e988.
  26. Nordmeyer J, Ewert P, Gewillig M, et al. Acute and midterm outcomes of the post-approval MELODY Registry: a multicentre registry of transcatheter pulmonary valve implantation. Eur Heart J 2019; 40:2255.
  27. Georgiev S, Ewert P, Tanase D, et al. A Low Residual Pressure Gradient Yields Excellent Long-Term Outcome After Percutaneous Pulmonary Valve Implantation. JACC Cardiovasc Interv 2019; 12:1594.
  28. Georgiev S, Ewert P, Eicken A, et al. Munich Comparative Study: Prospective Long-Term Outcome of the Transcatheter Melody Valve Versus Surgical Pulmonary Bioprosthesis With Up to 12 Years of Follow-Up. Circ Cardiovasc Interv 2020; 13:e008963.
  29. Jones TK, Rome JJ, Armstrong AK, et al. Transcatheter Pulmonary Valve Replacement Reduces Tricuspid Regurgitation in Patients With Right Ventricular Volume/Pressure Overload. J Am Coll Cardiol 2016; 68:1525.
  30. Tanase D, Ewert P, Georgiev S, et al. Tricuspid Regurgitation Does Not Impact Right Ventricular Remodeling After Percutaneous Pulmonary Valve Implantation. JACC Cardiovasc Interv 2017; 10:701.
  31. Kenny D, Rhodes JF, Fleming GA, et al. 3-Year Outcomes of the Edwards SAPIEN Transcatheter Heart Valve for Conduit Failure in the Pulmonary Position From the COMPASSION Multicenter Clinical Trial. JACC Cardiovasc Interv 2018; 11:1920.
  32. Hascoet S, Dalla Pozza R, Bentham J, et al. Early outcomes of percutaneous pulmonary valve implantation using the Edwards SAPIEN 3 transcatheter heart valve system. EuroIntervention 2019; 14:1378.
  33. Sinha S, Aboulhosn J, Asnes J, et al. Initial results from the off-label use of the SAPIEN S3 valve for percutaneous transcatheter pulmonary valve replacement: A multi-institutional experience. Catheter Cardiovasc Interv 2019; 93:455.
  34. Morray BH, McElhinney DB, Cheatham JP, et al. Risk of coronary artery compression among patients referred for transcatheter pulmonary valve implantation: a multicenter experience. Circ Cardiovasc Interv 2013; 6:535.
  35. Mauri L, Frigiola A, Butera G. Emergency surgery for extrinsic coronary compression after percutaneous pulmonary valve implantation. Cardiol Young 2013; 23:463.
  36. Boudjemline Y, Malekzadeh-Milani S, Patel M, et al. Predictors and outcomes of right ventricular outflow tract conduit rupture during percutaneous pulmonary valve implantation: a multicentre study. EuroIntervention 2016; 11:1053.
  37. McElhinney DB, Cheatham JP, Jones TK, et al. Stent fracture, valve dysfunction, and right ventricular outflow tract reintervention after transcatheter pulmonary valve implantation: patient-related and procedural risk factors in the US Melody Valve Trial. Circ Cardiovasc Interv 2011; 4:602.
  38. Nordmeyer J, Khambadkone S, Coats L, et al. Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation. Circulation 2007; 115:1392.
  39. Morgan GJ, Sadeghi S, Salem MM, et al. SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without "pre-stenting": A multi-institutional experience. Catheter Cardiovasc Interv 2019; 93:324.
  40. Buber J, Bergersen L, Lock JE, et al. Bloodstream infections occurring in patients with percutaneously implanted bioprosthetic pulmonary valve: a single-center experience. Circ Cardiovasc Interv 2013; 6:301.
  41. McElhinney DB, Benson LN, Eicken A, et al. Infective endocarditis after transcatheter pulmonary valve replacement using the Melody valve: combined results of 3 prospective North American and European studies. Circ Cardiovasc Interv 2013; 6:292.
  42. Cheung G, Vejlstrup N, Ihlemann N, et al. Infective endocarditis following percutaneous pulmonary valve replacement: diagnostic challenges and application of intra-cardiac echocardiography. Int J Cardiol 2013; 169:425.
  43. Dilber D, Hörer J, Malcic I, et al. Percutaneous pulmonary valve implantation and surgical valve replacement in patients with right ventricular outflow tract dysfunction--a complementary treatment concept. Int J Cardiol 2013; 169:e3.
  44. McElhinney DB, Sondergaard L, Armstrong AK, et al. Endocarditis After Transcatheter Pulmonary Valve Replacement. J Am Coll Cardiol 2018; 72:2717.
  45. Corno AF, Hurni M, Griffin H, et al. Bovine jugular vein as right ventricle-to-pulmonary artery valved conduit. J Heart Valve Dis 2002; 11:242.
  46. Breymann T, Boethig D, Goerg R, Thies WR. The Contegra bovine valved jugular vein conduit for pediatric RVOT reconstruction: 4 years experience with 108 patients. J Card Surg 2004; 19:426.
  47. Bové T, Demanet H, Wauthy P, et al. Early results of valved bovine jugular vein conduit versus bicuspid homograft for right ventricular outflow tract reconstruction. Ann Thorac Surg 2002; 74:536.
  48. Carrel T, Berdat P, Pavlovic M, Pfammatter JP. The bovine jugular vein: a totally integrated valved conduit to repair the right ventricular outflow. J Heart Valve Dis 2002; 11:552.
  49. Boudjemline Y, Bonnet D, Massih TA, et al. Use of bovine jugular vein to reconstruct the right ventricular outflow tract: early results. J Thorac Cardiovasc Surg 2003; 126:490.
  50. Tiete AR, Sachweh JS, Roemer U, et al. Right ventricular outflow tract reconstruction with the Contegra bovine jugular vein conduit: a word of caution. Ann Thorac Surg 2004; 77:2151.
  51. Meyns B, Van Garsse L, Boshoff D, et al. The Contegra conduit in the right ventricular outflow tract induces supravalvular stenosis. J Thorac Cardiovasc Surg 2004; 128:834.
  52. Shebani SO, McGuirk S, Baghai M, et al. Right ventricular outflow tract reconstruction using Contegra valved conduit: natural history and conduit performance under pressure. Eur J Cardiothorac Surg 2006; 29:397.
  53. Van Dijck I, Budts W, Cools B, et al. Infective endocarditis of a transcatheter pulmonary valve in comparison with surgical implants. Heart 2015; 101:788.
  54. Hascoet S, Mauri L, Claude C, et al. Infective Endocarditis Risk After Percutaneous Pulmonary Valve Implantation With the Melody and Sapien Valves. JACC Cardiovasc Interv 2017; 10:510.
  55. Miranda WR, Connolly HM, Bonnichsen CR, et al. Prosthetic pulmonary valve and pulmonary conduit endocarditis: clinical, microbiological and echocardiographic features in adults. Eur Heart J Cardiovasc Imaging 2016; 17:936.
  56. Bergersen L, Benson LN, Gillespie MJ, et al. Harmony Feasibility Trial: Acute and Short-Term Outcomes With a Self-Expanding Transcatheter Pulmonary Valve. JACC Cardiovasc Interv 2017; 10:1763.
  57. Morgan GJ, Sivakumar K, Promphan W, et al. Early clinical experience with the straight design of Venus P-valve™ in dysfunctional right ventricular outflow tracts. Catheter Cardiovasc Interv 2020; 96:E653.
  58. Kim GB, Song MK, Bae EJ, et al. Successful Feasibility Human Trial of a New Self-Expandable Percutaneous Pulmonary Valve (Pulsta Valve) Implantation Using Knitted Nitinol Wire Backbone and Trileaflet α-Gal-Free Porcine Pericardial Valve in the Native Right Ventricular Outflow Tract. Circ Cardiovasc Interv 2018; 11:e006494.
  59. Zahn EM, Chang JC, Armer D, Garg R. First human implant of the Alterra Adaptive PrestentTM : A new self-expanding device designed to remodel the right ventricular outflow tract. Catheter Cardiovasc Interv 2018; 91:1125.
Topic 86481 Version 18.0

References

1 : Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction.

2 : Extended application of percutaneous pulmonary valve implantation.

3 : Outcomes and safety of transcatheter pulmonary valve replacement in patients with large patched right ventricular outflow tracts.

4 : Melody valve implant within failed bioprosthetic valves in the pulmonary position: a multicenter experience.

5 : Cracking the ring of Edwards Perimount bioprosthesis with ultrahigh pressure balloons prior to transcatheter valve in valve implantation.

6 : Percutaneous pulmonary valve implantation: the Munich experience

7 : Percutaneous pulmonary valve implantation: two-centre experience with more than 100 patients.

8 : Short- and medium-term outcomes after transcatheter pulmonary valve placement in the expanded multicenter US melody valve trial.

9 : ESC Guidelines for the management of grown-up congenital heart disease (new version 2010).

10 : Images in cardiovascular medicine. Transcatheter right ventricular outflow tract intervention: the risk to the coronary circulation.

11 : Early clinical experience with a novel self-expanding percutaneous stent-valve in the native right ventricular outflow tract.

12 : Early European experience with the Venus P-valve®: filling the gap in percutaneous pulmonary valve implantation.

13 : Implantation of the melody transcatheter pulmonary valve in patients with a dysfunctional right ventricular outflow tract conduit early results from the u.s. Clinical trial.

14 : Percutaneous replacement of pulmonary valve using the Edwards-Cribier percutaneous heart valve: first report in a human patient.

15 : Transcatheter pulmonary valve implantation using the Edwards SAPIEN transcatheter heart valve.

16 : Percutaneous implantation of the Edwards SAPIEN transcatheter heart valve for conduit failure in the pulmonary position: early phase 1 results from an international multicenter clinical trial.

17 : Pre-stenting with a bare metal stent before percutaneous pulmonary valve implantation: acute and 1-year outcomes.

18 : Tricuspid regurgitation as a complication of Edwards Sapien XT valve implantation in pulmonary position a problem to deal with.

19 : Use of 65 cm large caliber Dryseal sheaths to facilitate delivery of the Edwards SAPIEN valve to dysfunctional right ventricular outflow tracts.

20 : Percutaneous pulmonary valve-in-valve implantation: a successful treatment concept for early device failure.

21 : Transvenous, antegrade Melody valve-in-valve implantation for bioprosthetic mitral and tricuspid valve dysfunction: a case series in children and adults.

22 : Percutaneous pulmonary valve implantation: impact of evolving technology and learning curve on clinical outcome.

23 : Melody transcatheter pulmonary valve implantation. Results from the registry of the Italian Society of Pediatric Cardiology.

24 : Clinical and hemodynamic outcomes up to 7 years after transcatheter pulmonary valve replacement in the US melody valve investigational device exemption trial.

25 : Up to 11 years of experience with the Melody valved stent in the right ventricular outflow tract.

26 : Acute and midterm outcomes of the post-approval MELODY Registry: a multicentre registry of transcatheter pulmonary valve implantation.

27 : A Low Residual Pressure Gradient Yields Excellent Long-Term Outcome After Percutaneous Pulmonary Valve Implantation.

28 : Munich Comparative Study: Prospective Long-Term Outcome of the Transcatheter Melody Valve Versus Surgical Pulmonary Bioprosthesis With Up to 12 Years of Follow-Up.

29 : Transcatheter Pulmonary Valve Replacement Reduces Tricuspid Regurgitation in Patients With Right Ventricular Volume/Pressure Overload.

30 : Tricuspid Regurgitation Does Not Impact Right Ventricular Remodeling After Percutaneous Pulmonary Valve Implantation.

31 : 3-Year Outcomes of the Edwards SAPIEN Transcatheter Heart Valve for Conduit Failure in the Pulmonary Position From the COMPASSION Multicenter Clinical Trial.

32 : Early outcomes of percutaneous pulmonary valve implantation using the Edwards SAPIEN 3 transcatheter heart valve system.

33 : Initial results from the off-label use of the SAPIEN S3 valve for percutaneous transcatheter pulmonary valve replacement: A multi-institutional experience.

34 : Risk of coronary artery compression among patients referred for transcatheter pulmonary valve implantation: a multicenter experience.

35 : Emergency surgery for extrinsic coronary compression after percutaneous pulmonary valve implantation.

36 : Predictors and outcomes of right ventricular outflow tract conduit rupture during percutaneous pulmonary valve implantation: a multicentre study.

37 : Stent fracture, valve dysfunction, and right ventricular outflow tract reintervention after transcatheter pulmonary valve implantation: patient-related and procedural risk factors in the US Melody Valve Trial.

38 : Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation.

39 : SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without "pre-stenting": A multi-institutional experience.

40 : Bloodstream infections occurring in patients with percutaneously implanted bioprosthetic pulmonary valve: a single-center experience.

41 : Infective endocarditis after transcatheter pulmonary valve replacement using the Melody valve: combined results of 3 prospective North American and European studies.

42 : Infective endocarditis following percutaneous pulmonary valve replacement: diagnostic challenges and application of intra-cardiac echocardiography.

43 : Percutaneous pulmonary valve implantation and surgical valve replacement in patients with right ventricular outflow tract dysfunction--a complementary treatment concept.

44 : Endocarditis After Transcatheter Pulmonary Valve Replacement.

45 : Bovine jugular vein as right ventricle-to-pulmonary artery valved conduit.

46 : The Contegra bovine valved jugular vein conduit for pediatric RVOT reconstruction: 4 years experience with 108 patients.

47 : Early results of valved bovine jugular vein conduit versus bicuspid homograft for right ventricular outflow tract reconstruction.

48 : The bovine jugular vein: a totally integrated valved conduit to repair the right ventricular outflow.

49 : Use of bovine jugular vein to reconstruct the right ventricular outflow tract: early results.

50 : Right ventricular outflow tract reconstruction with the Contegra bovine jugular vein conduit: a word of caution.

51 : The Contegra conduit in the right ventricular outflow tract induces supravalvular stenosis.

52 : Right ventricular outflow tract reconstruction using Contegra valved conduit: natural history and conduit performance under pressure.

53 : Infective endocarditis of a transcatheter pulmonary valve in comparison with surgical implants.

54 : Infective Endocarditis Risk After Percutaneous Pulmonary Valve Implantation With the Melody and Sapien Valves.

55 : Prosthetic pulmonary valve and pulmonary conduit endocarditis: clinical, microbiological and echocardiographic features in adults.

56 : Harmony Feasibility Trial: Acute and Short-Term Outcomes With a Self-Expanding Transcatheter Pulmonary Valve.

57 : Early clinical experience with the straight design of Venus P-valve™in dysfunctional right ventricular outflow tracts.

58 : Successful Feasibility Human Trial of a New Self-Expandable Percutaneous Pulmonary Valve (Pulsta Valve) Implantation Using Knitted Nitinol Wire Backbone and Trileafletα-Gal-Free Porcine Pericardial Valve in the Native Right Ventricular Outflow Tract.

59 : First human implant of the Alterra Adaptive PrestentTM : A new self-expanding device designed to remodel the right ventricular outflow tract.